cobas® pulse is a point-of-care blood glucose (BG) measuring system for multiple-patient use in professional healthcare settings. The system provides advances in connectivity and BG measuring technology, and has multiple fail-safes to improve accuracy and reduce the risk of user error. Flavin adenine dinucleotide-dependent glucose dehydrogenase on the working electrode catalyzes oxidation of β-D-glucose in the blood sample. A redox mediator/electron acceptor, on both the working and the counter electrode, facilitates diffusion of electrons in proportion to the glucose concentration and compensates for the effects of potential interfering agents. During development, >1 million test strip measurements were performed using >8000 test scenarios to refine the algorithm model. No clinically relevant interference was identified with extreme variations in blood properties and drugs in whole blood samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.