Abstract

We studied coaxial dual flow ventilation circuit which was combined with the humidifier controlling inspired gas temperature automatically.The variation of inspired gas temperature was controlled within ±0.3°C when tidal volume and respiratory rate were changed. Inspired gas temperature indicated stably within ±0.2°C under the change of specified temperature of inspired gas and ventilation. In addition, inspired gas temperature was little influenced by the change of room temperature.Rainout in the circuit slightly increased with decreasing of respiratory rate, however, its value was lower than that of conventional circuit, as this circuit is dual flow system warming an inspired limb with heat of expired gas.From the above results, we conclude that this system can control inspired gas temperature accurately and simply, and there is less rainout in the circuit.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.