Renal blood flow (RBF) may vary during cardiopulmonary bypass and low flow may cause insufficient blood supply of the kidney triggering renal failure postoperatively. Still, a valid intraoperative method of continuous RBF measurement is not available. A new catheter combining thermodilution and intravascular Doppler was developed, first calibrated in an in vitro model, and the catheter specific constant was determined. Then, application of the device was evaluated in a pilot study in an adult cardiovascular population. The data of the clinical pilot study revealed high correlation between the flow velocities detected by intravascular Doppler and the RBF measured by thermodilution (Pearson's correlation range: 0.78 to 0.97). In conclusion, the RBF can be measured excellently in real time using the new catheter, even under cardiopulmonary bypass.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.