Abstract

Considering the damage to normal tissues during the high-intensity focused ultrasound ablation of the tumor is important. The purpose of this work was to design the annular array under the considerations of preventing the skin burn and damage to normal tissues behind the tumor target. Based on the Rayleigh-Summerfield pressure integral, the numerical acoustic intensity field of a spherical bowl annular array with a diameter of 12cm and a radius of curvature of 12cm was obtained by using the MATLAB software. An absorbed intensity ratio of the normal tissue to target was proposed to define the allowable grating lobe for determining the focusing range. The analytic results showed that for the treatment of uterine fibroids, the focusing range increased as the number of the element increased from 9 to 22 and the 14-ring array at the operating frequency of 0.8, 0.9 or 1.0MHz had a focusing range larger than 5.5cm when the range was only 4cm at 1.2MHz. For the treatment of medullary carcinoma and scirrhous carcinoma in breast, the focusing range of the 0.9-MHz, 14-ring annular array was 8-12 and 8-14cm, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.