Abstract

The aim of this experimental study was to evaluate a newly designed cone-shaped aortic valve prosthesis (CAVP) for one-step transcatheter placement in an orthotopic position. The study was conducted in 15 swine using either the transcarotid (11 animals) or the transfemoral (4 animals) artery approach. A 12- or 13-Fr sheath was inserted via arterial cutdown. The CAVP was deployed under fluoroscopic control and its struts, by design, induced significant native valve insufficiency. CAVP function was evaluated by aortography and aortic pressure curve tracing. In 11 of 15 swine the CAVP was properly deployed and functioned well throughout the scheduled period of 2-3 h. In three swine the CAVPs were placed lower than intended, however, they were functional even in the left ventricular outflow tract position. One swine expired due to inadvertent low CAVP placement that caused both aortic regurgitation and immobilization of the anterior mitral valve leaflet by the valve struts. We conclude that this design of CAVP is relatively easy to deploy, works well throughout a short time period (2-3 h), and, moreover, seems to be reliable even in a lower-than-orthotopic position (e.g., infra-annulary space). Longer-term studies are needed for its further evaluation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.