Abstract

A compact 60 kV Mott polarimeter has been constructed, tested, and calibrated especially for the local analysis of surface and two-dimension magnetism by means of electron spectroscopy with spin resolution. The use of construction combining spherical accelerating field with the absence of retarding potential after electron scattering provides low sensitivity of the polarimeter to movements of the analyzed electron beam and changes in its diameter. Special silicon surface-barrier detectors with large electron-optical acceptance provide efficiency ∈(≡Seff2×I/I0) about ≈2.5×10−4. The unique fast electronics allow to work with maximum count rate of detected electrons up to 5×106 counts/s. As a result, there is the possibility to calibrate the polarimeter (to calculate the effective Sherman function Seff) by extrapolation of the measured asymmetry to the high level of discrimination. This instrument is compact, fully UHV compatible and may be used in other fields of solid state physics, atomic physics, and physics of high energies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.