Abstract

The FI (partially frozen injectate) system, a new closed-system devised by the authors for thermodilution cardiac output determinations, has two major features: 1) it needs no ice-filled receptacle to keep injectate cold because it uses partially frozen injectate, and 2) it can go without monitoring the injectate temperatures during the whole process of cardiac output determinations. The author evaluated the accuracy and reproducibility of cardiac output determinations with the FI system in 10 critically ill patients, as compared with another closed-system (which is commercially available) and the standard open method. The injectate temperatures in the FI system were also measured in vitro. The mean injectate temperature in the FI system was 0.71 +/- 0.26 degrees C and 80% of the injectate temperatures were lower than 1.0 degrees C. Even when no monitoring of injectate temperatures was made, the predicated error in the calculated cardiac output resulted as low as 2% with the FI system. The mean cardiac output values were not statistically different between the FI system and the other two systems.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.