Abstract
The syringe infusion pump has been established as the instrument of choice for high‐risk infusions, where potent drugs are often delivered at low rates of flow. However, numerous instances of unexpected flow error with consequent patient physiological impact have been reported. These include unwanted bolus delivery on release of line occlusion, dosage fluctuation due to pump height change and fluid reflux within the multiple pump installations now common in the Intensive Care Unit (ICU). This article examines the performance of a typical ICU syringe infusion pump and identifies mechanical compliance, inherent in commercial designs, as a source of flow error that should not be ignored by equipment designers. A prototype low compliance system is described and tested with performance compared to the conventional design, demonstrating advantages in terms of lower flow error.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.