Abstract

This work deals with the determination of fluorine by solid sampling graphite furnace molecular absorption spectrometry. The molecular absorbance of aluminum monofluoride (AlF), which is produced in the vapor phase in the presence of Al 3+, is measured at 227.5 nm, a non-resonant platinum line. A conventional graphite furnace program has been used with pyrolysis and vaporization temperatures of 800 and 2300 °C, respectively. Solutions of Ba 2+ and Al 3+ have been used to avoid fluorine losses during the pyrolysis stage and to produce AlF in the vaporization stage, respectively. Certified coal and alumina samples were analyzed using aqueous standards for calibration. The agreement between the found concentration and the certified value, or the value obtained by another method ranged from 92 to 105%, with a relative standard deviation less than 8.5%. The limit of detection and the characteristic mass was 0.17 μg g − 1 and 205 pg, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.