Abstract

No studies have compared the accuracy of acetabular cup placement in total hip arthroplasty (THA) in the supine and lateral decubitus positions using the same portable navigation system. Thus, this study aimed to compare the accuracy of acetabular cup placement using a new accelerometer-based portable navigation system combined with an infrared stereo camera and inertial measurement unit between the supine and lateral decubitus positions. This retrospective study compared 45 THAs performed in the supine position (supine group) and 44 THAs performed in the lateral decubitus position (lateral group) using the same portable navigation system. The primary outcome was the absolute errors of cup placement angles, defined as the absolute values of the differences between cup radiographic inclination and anteversion angles displayed on the navigation system and those measured on postoperative computed tomography images. No significant difference in the median absolute error of the cup inclination angle (supine group 1.7° [interquartile range 0.8°-3.1°] vs. lateral group 2.1° [interquartile range 1.0°-3.7°]; p = 0.07) was found between the two groups. Similarly, no significant difference in the median absolute error of the anteversion angle (supine group 1.9° [interquartile range 0.8°-3.4°] vs. lateral group 2.1° [interquartile range 0.9°-3.1°]; p = 0.42) was found. This new accelerometer-based portable navigation system may provide high accuracy of the cup placement in THA in the lateral decubitus and supine positions.