Abstract

Microchannel plates containing neutron absorbing elements such as boron and gadolinium in the bulk glass are used as the sensing element in high spatial resolution, high rate neutron imaging systems. In this paper we describe one such device, using both <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">10</sup> B and natural Gd, which employs cross delay line signal readout, with time-of-flight capability. This detector has a measured spatial resolution under 40 μm FWHM, thermal neutron efficiency of 19%, and has recorded rates in excess of 500 kHz. A physical and functional description is presented, followed by a discussion of measurements of detector performance and a brief survey of some practical applications.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call