Abstract
A neutron guide system that includes neutron guides, a main shutter, and a vacuum system was successfully installed at the HANARO research reactor of the Korea Atomic Energy Research Institute (KAERI) last year, and is now operating with 5 cold neutron instruments. The neutron flux and spectrum were measured by using gold wire and a disc chopper. The total measured neutron fluxes for various position are about 10–25% lower than the calculated fluxes, which is probably caused by neutron guide misalignment, larger gap between neutron guides, low reflectivity, imperfect cold neutron source data, and so on. But the measured neutron fluxes of the neutron guides are very high. The status of the neutron guide installation and its first performance test result is described in this paper.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.