Coriolis mass flow meters provide accurate measurement of single-phase flows, typically to 0.2%. However gas–liquid two-phase flow regimes may cause severe operating difficulties as well as measurement errors in these flow meters. As part of the Sensor Validation (SEVA) research at Oxford University a new fully digital coriolis transmitter has been developed which can operate with highly aerated fluids. This paper describes how a neural network has been used to correct the mass flow measurement for two-phase flow effects, based entirely on internally observed parameters, keeping errors to within 2%. The correction strategy has been successfully implemented on-line in the coriolis transmitter. As required by the SEVA philosophy, the quality of the corrected measurement is indicated by the on-line uncertainty provided with each measurement value.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.