Abstract

Real-time glucose monitoring has been beneficial in reducing health complications associated with diabetes as well as a decrease in mortality. This report describes a novel holographic platform, fabricated via laser ablation on chitosan hydrogel with gold nanoparticles with a replaying in visible and near IR. The sensor responded with a 12nm and 7nm shift in wavelength at glucose concentrations in the 0–70mM range and in the visible and near IR, respectively, at pH 7.4 and an ionic strength of 154mM. The sensor did not respond to potential interferences found in the interstitial fluid, such as fructose, vitamin C and lactate, at their respective normal concentrations and was stable to fluctuations in temperature, pH and ionic strength. The characteristics of this sensor suggests that it may be applicable for use as an implanted device for the real time monitoring of glucose concentrations in the interstitial fluid using near IR as the interrogating medium.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.