Abstract

An innovative high-performance liquid chromatography assay method was developed and validated for quantification of dextromethorphan hydrobromide and desloratadine simultaneously in monophasic liquid formulation by preparing syrup containing 30mg/5mL of dextromethorphan hydrobromide and 1.2mg/mL of desloratadine. The chromatographic severance was executed by gradient solution A and B. The composition of buffer solution A contained 0.05M monobasic potassium, then 1mL triethylamine was added to it and the pH was adjusted to 2.3 with orthophosphoric acid. Methanol was used as solution B. The gradient elution was executed with Kromasil C8 (250mm×4.6mm) column having 1.5mL/min flow rate and 20µL injection volume with UV-estimation at 254nm for dextromethorphan hydrobromide and DES. The present research was planned according to Box-Behnken design by utilizing design expert software, using four factors such as column temperature (A), flow rate (B), mobile phase-organic phase (C), and pH (D); correspondingly the selected response variables were resolution between A and B, that is, desloratadine and methyl paraben (Y1), tailing of dextromethorphan hydrobromide (Y2), and tailing of desloratadine (Y3). The parameters such as system suitability, linearity, accuracy, precision, robustness, limit of detection, limit of quantitation, and ruggedness were analyzed to validate the developed method in accordance with current regulatory guidelines.