Abstract

A new imaging sequence for rapid determination of the apparent self-diffusion tensor of water was developed and tested on fixed excised rat spinal cords. To reduce the time required to determine the tensor, the sequence utilized a new single-shot approach with multiple spin echoes. An assumption of cylindrical symmetry in the sample was made, thus requiring the measurement of only four of the six unique elements of the tensor. This assumption was found experimentally to be valid, and the results obtained using the new sequence were found to be quantitatively the same as results obtained using a standard spin-echo sequence.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.