Abstract

This paper presents a multimodal multichannel neural activity readout IC which can perform not only the electrical recording (ER) but also the fluorescence recording (FR) of neural activity for the cell-type-specific study of heterogeneous neuronal cell populations. The time-based FR circuit senses Ca2+ concentration using Ca2+ probes while the ER circuit acquires action potentials (APs) and local field potentials (LFPs). The IC is fabricated in 0.18μm CMOS. The FR circuit achieves a recording range of 81dB (75pA to 860nA) and consumes the power of 0.7μmW/Ch. The ER circuit achieves the input-referred noise (IRN) of 2.7μm V rms over the bandwidth (BW) of 10kHz, while consuming the power of 4.9μmW/Ch. The in-vitro measurement is performed for recording Ca2+ concentration and electrical neural signals.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.