Abstract

ABSTRACTThe successful modification of the tips of a cellular microgripper into ion selective electrodes capable of sensing calcium ions at concentrations as low as 8x10-5 M is described. The modification involves applying the process of adding the components of all solid state ion selective electrodes. Specifically, poly(3,4-ethylenedioxythiophene) (PEDOT) is added to a gold electrode protruding from the microgripper tip; this is then coated with a poly(vinyl chloride) PVC based calcium selective membrane. Excellent Nernstian response was observed from our devices, with calibration slopes of 29.5 ± 2.5 mV/dec.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.