Abstract

In this work, a flow analysis procedure for the determination of copper, chromium, iron and lead in lubricating oils using flame AAS as detection technique is described. The flow manifold was designed to implement the multicommutation approach and it comprised three 3-way solenoid valves controlled by a personal computer. The flow system presented allowed to process the oil samples to determine wear metals without any prior preparation. Aiming to assess accuracy the results were compared with those obtained by manual procedure using flame AAS. Applying the joint-confidence ellipse test, no significant difference at the 95% confidence level was observed. Other profitable features such as a sample throughput of 50 determinations per hour; relative standard deviations ( n = 5) below 2% for Cu, and below 8% for Cr, Fe and Pb; and linear responses in the range 0–40 ppm (w/w) (Cu, Fe) and 0–15 ppm (w/w) (Cr, Pb) were also achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.