Abstract

We have built a high-resolution high-efficiency superconducting soft x-ray spectrometer for synchrotron-based fluorescence-detected absorption spectroscopy. The sensor is a 3×3 array of 200×200 μm2 superconducting Nb–Al–Al2O3–Al–Nb tunnel junctions with an energy resolution around 15 eV below 1 keV and a total count rate capability of ≈100 000 counts/s. This sensor array is cooled to ≈0.1 K by a two-stage adiabatic demagnetization refrigerator while held at the end of a 40-cm-long cold finger that can be inserted into an UHV sample chamber for x-ray fluorescence measurements. We present L-edge absorption spectra of dilute transition metals (≈few 100 ppm) and discuss spectrometer performance with respect to the analysis of metalloproteins.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.