Abstract
The design of telemetry systems for chronic implantation within the body have progressed from simple single-channel devices to complex multichannel systems over the last 15 yr. Although chronic measurements of temperature, ECG, and pressure have been reported with good results, measurements such as dimension and blood flow have been difficult because of heavy power requirements. The design to be described is a multiplex system that will measure up to eight individual parameters simultaneously, including blood flow. Flow is measured using an electromagnetic transducer, and by special design, the normal high power requirements of the electromagnetic technique have been reduced to a few hundred milliwatts. The system is powered by two NiCad, rechargeable batteries which are periodically recharged through the intact skin by induction at 250 kHz to an implanted pickup coil. The presently constructed units have been configured to measure ECG, two pressures, temperature and ascending aortic flow.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.