Abstract

We have developed a multi-channel electrochemical immunosensing system for the detection of α-fetoprotein(AFP).The sensing system consisted of carbon screen-printed electrode(SPCE) arrays and poly HRP-based signal amplification probes.AFP antigens could bind to both capture antibodies immobilized on electrode arrays and detection antibodies(anti-AFP Ig G developed in rabbit). The anti-rabbit Ig G conjugated to poly-horseradish peroxidase were used as a signaling probe to bind to the immuno-complex. The current signals were harvested by a home-made multi-channel electrochemical detector. AFP could be detected in the concentration ranges of 64 pg·m L-1~ 250 ng·m L-1. The detection limit was as low as 56 pg·m L-1. The proposed immunosensing system provided a high sensitive, specific, simple and practical method for AFP detection.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.