Abstract

All-solid-state potentiometric sensors that are easily miniaturized and arrayed are widely used in home diagnostics. However, changes in sample matrix compositions tend to affect the basic potential of the potentiometric sensor, and pH of sample could change the response slope, thus affecting the detection reliability. This study takes the detection of glucose in urine as a model to increase the reliability of potentiometric sensors in home detection. PtAu/CNTs nanozyme modified by diboronic acid has been designed, showing better catalytic selectivity for glucose by experiments and theoretical calculations. Moreover, glucose electrode group in a multi-calibration glucose potentiometric sensing array can realize the basic potential calibration of sensing channel by the calibration channel. Meanwhile, the pH electrode group can not only measure the urine pH, but also calibrate the response slope of the glucose electrode group, thus improving the reliability of home detection.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call