Abstract

Photons and protons are usually used independently to treat cancer. However, at MGH patients can be treated with both photons and protons since both modalities are available on site. A combined therapy can be advantageous in cancer therapy due to the skin sparing ability of photons and the sharp Bragg peak fall-off for protons beyond the tumor. In the present work, we demonstrate how to implement a combined 3D MC toolkit for photon and proton (ph-pr) therapy, which can be used for verification of the treatment plan. The commissioning of a MC system for combined ph-pr involves initially the development of a MC model of both the photon and proton treatment heads. The MC dose tool was evaluated on a head and neck patient treated with both combined photon and proton beams. The combined ph-pr dose agreed with measurements in solid water phantom to within 3%mm. Comparison with commercial planning system pencil beam prediction agrees within 3% (except for air cavities and bone regions).