Abstract
For an optimal tuning of the radiation protection planning, a Monte Carlo model using the MCNPX code has been built, allowing an accurate estimate of the spectrometric and geometrical characteristics of photoneutrons generated by a Varian TrueBeam Stx© medical linear accelerator. We considered in our study a device working at the reference energy for clinical applications of 15MV, stemmed from a Varian Clinac©2100 modeled starting from data collected thanks to several papers available in the literature. The model results were compared with neutron and photon dose measurements inside and outside the bunker hosting the accelerator obtaining a complete dose map. Normalized neutron fluences were tallied in different positions at the patient plane and at different depths. A sensitivity analysis with respect to the flattening filter material were performed to enlighten aspects that could influence the photoneutron production.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.