Abstract
Due to the very high complexity of IMRT treatment plans, it is imperative to perform dose verification, preferably before patient delivery. The aim of this project is to develop a Monte-Carlo-based model to verify the final dose distributions of plans developed using the Peacock system (CORVUS Treatment Planning System and MIMiC collimator). The system delivers radiation through arc therapy and uses sinogram files to determine the state of each of the multileaf collimator leaves. In-house software was developed using Matlab to decode the sinograms and create blocklets that are used as input in an MCSIM model of the MIMiC collimator attached to a Varian Clinac 600C. After validating the model, a prostate and head and neck case were simulated. The CORVUS-predicted dose distributions were compared with the Monte Carlo dose distributions. As expected, the results agreed very closely for the homogeneous case of the prostate but there were large discrepancies observed for the more heterogeneous head and neck case.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.