Abstract

This work shows the effectiveness of Monte Carlo methods for calibrating a dual-energy CT (DECT) system on a heterogeneous phantom for radiotherapy. The reference phantom comprising 20 inserts of different materials and densities representing human tissues inserted in a plastic water cylinder has been considered. The reliability of the model, built with the Penelope code, has been verified by comparing the results of the simulations at 80[Formula: see text]kVp and 140[Formula: see text]kVp with the experimental data acquired from the CT scans and dosimetry measurements. The effective atomic number from the various materials is also considered a control parameter. Tests for calibration and dose calculation are also based on a homogeneous phantom in PMMA and measurements in air with an ionization chamber. The Monte Carlo simulations coupled with dose evaluations allowed the calibration of the dose deposition effects on the considered tissue models.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.