Abstract
A fully integrated SoC for multimodality intracranial neuromonitoring is presented in this paper. Three sensors including a capacitive MEMS pressure sensor, an electrochemical oxygen sensor and a solid-state temperature sensor are integrated together in a single chip with their respective interface circuits. Chopper stabilization and dynamic element matching techniques are applied in sensor interface circuits to reduce circuit noise and offset. On-chip calibration is implemented for each sensor to compensate process variations. Measured sensitivity of the pressure, oxygen, and temperature sensors are 18.6 aF/mmHg, 194 pA/mmHg, and 2 mV/°C, respectively. Implemented in 0.18 m CMOS, the SoC occupies an area of 1.4 mm × 4 mm and consumes 166 μW DC power. A prototype catheter for intracranial pressure (ICP) monitoring has been implemented and the performance has been verified with ex vivo experiment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.