Abstract

To meet the needs in gas-sweetening processes for on-line monitoring of low levels of hydrogen sulfide in alkanolamine solutions, an on-line analysis system has been developed based on the measurement of the intense and broad ultraviolet (u.v.) absorption of the H 2S-amine bond. This system incorporates the following three features to provide the requird accuracy and reliability. To compensate for variation in sample background absorbance from additives and from degradation and corrosion products, a vacuum distillation step is used to strip out H 2S from a process sample to provide an H 2S-free “zero” standard. To compensate for the temperature sensitivity of the u.v. absorption of the H 2S-amine bond, close temperature control as well as a temperature-compensating circuit is provided. To handle sample streams often containing heavy loadings of fine and well-dispersed particulate matter, a special self-cleaning filtering system is used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.