Abstract

An monitor for control of accelerated particle flux density used in irradiation of polymer films with heavy ions with a low track density (103−5 × 107 cm−2) is based on detection of recoil protons. It has been designed to control irradiation of a film with a width of up to 650 mm and is composed of five PIN diodes with a hydrogenous target placed in front of each diode. The sensitivity of the monitor with a polyethylene target to the density of a xenon ion beam is described by a linear function; its value is (4.0±0.5) × 10−3 pulses/(ion/cm2). The instrument is capable of monitoring ion flux densities of <106 cm−2, for which all other methods are ineffective. It can be used to determine the track density in a finished product, bypassing the stage of a check with an electron microscope. The signals from the monitor can also be used to introduce a feedback in a system of automatic beam tuning in order to improve the film irradiation quality. The radiation hardness of silicon detectors is ≈ 1012 protons/cm2. At an average detector load of ≈ 104 protons/cm2, the average service life of the monitor is 108 s of continuous operation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.