Abstract
AbstractBeam brightness modulation Auger spectrometry is shown to be an effective means of eliminating ion induced Auger electrons from Auger electron energy spectra. Modulating the incident electron beam intensity causes the true electron induced Auger electron emission to be modulated at the beam modulation frequency which can be easily separated from the unmodulated ion induced Auger electrons by means of frequency sensitive detection using a lock‐in amplifier. This technique is particularly useful when depth profiling with a high energy ion beam (∼5 kV) and a low current (small spot) electron beam (<100 nA).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.