Abstract

We present a new tool for the precisely controlled transfer of individual picoliter (pL) droplets in the range of 150-950 pL at user defined local positions within aqueous liquid environments while avoiding any leakage by diffusion. This is achieved by a low-cost, disposable and biocompatible cap that can be placed on top of any pL-dispenser and generates a phase-gap between dispensing agent and target liquid when the dispenser is dipped into the latter. We developed two different working modes: (i) the standard mode enables an instant injection (<< 1 ms) of the droplet into the liquid environment and (ii) the focus mode further increases the spatial resolution from 100 microm to 50 microm at the cost of slowing down the injection time. For the phase-gap we have proven an excellent long-term stability of more than 30 hours against capillary priming.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.