Although manual enumeration of Chlamydia inclusion forming units is the most widely accepted means of quantification in the field, it is both time consuming and subject to inherent investigator bias. We report here a rapid, i.e., minutes vs. hours, modified automated Fluorospot means of assessment that is linear (<1200 dots per well). Because the Fluorospot enumerated tissue culture plate/well can also be quantified using traditional manual counting, newly derived Fluorospot data can easily be compared to previously established manual enumeration data requiring no new reference norms.•Concurrent enumeration of chlamydial IFU using automated and manual methods of counting on same tissue culture plate.•Rapid method of counting chlamydial IFU reducing time from hours to minutes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.