Abstract

The HeartWare left ventricular assist device is a miniaturized, continuous centrifugal-flow pump. The implantation technique is well described and relatively standardized across different institutions. However, there still exists a technical concern about handling the inflow cannula at the time of device explant. Specifically, the removal of the sewing ring and plicating the apical defect en masse may distort the geometry of the left ventricle and impart myocardial dysfunction. Additionally, a prefabricated repair mechanism by the manufacturer is not readily available in all countries (i.e. USA). Here, we describe a technique to address the apical core defect, using a tailor-made plug and leaving the sewing ring in situ, at the time of the HeartWare left ventricular assist device explant.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.