Abstract
AbstractThe widely used formalism for quantitative XRF analysis based on the correction of characteristic line intensity for absorption effects is shown to be inadequate when thick samples with low total mass attenuation coefficients are analysed (biological samples, solutions, etc.). The geometry factor is not a constant function of the collimator's shape and dimensions, but it is found that in an analysed matrix it depends on the line energy and penetration depth. An experimental and a theoretical model for the determination of the geometrical correction are proposed and tested on a set of samples. The concentrations obtained are in excellent agreement with the values expected, whereas the model based on the correction of intensities only for absorption effects gives an error of at least 25% for thick samples.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.