Abstract

Abstract In this paper we propose a model-based scheme to control the depth of hypnosis in anesthesia that uses the BIS signal as controlled variable. In particular, the control scheme exploits the propofol pharmacokinetics/pharmacodynamics model of the patient so that the estimated effect-site concentration is used as a feedback signal for a standard PID controller, which compensates for the model uncertainties. The tuning of the parameters is performed off-line using genetic algorithms to minimize a performance index over a given data set of patients. The effectiveness of the proposed method is verified by means of a Monte Carlo method that takes into account both the intra-patient and inter-patient variability. In general, we obtain a fast induction phase with limited overshoot and a good disturbance rejection during maintenance of anesthesia.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.