Abstract

In order for vacuum-assisted venous return (VAVR) to be used safely and efficiently during cardiopulmonary bypass (CPB), a full understanding of venous return is necessary. The focus of this work was to use the concepts of energy conservation and viscous energy dissipation in the development of a theoretical model of venous return utilizing vacuum assist. The effectiveness and accuracy of this model has been verified through in vitro laboratory investigations and statistical analysis. Although VAVR can provide higher flows through smaller venous cannula, vacuum assist may lead to increased levels of wall shear stress as shown in this work. The clinical implications of VAVR have yet to be investigated, but may lead to an exacerbation of the detrimental effects of CPB during cardiac surgery.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.