Abstract

A model of the behavior of spins moving through spatially varying gradient and B1 fields is presented. The model simulates the adiabatic behavior of flowing arterial water during a two-coil arterial inversion experiment. Predictions of the degree of inversion generated by the model are compared with flow phantom results for a wide range of gradient magnitudes, nominal B1 magnitudes, and flow velocities. The high level of agreement between the model and the flow phantom results indicates that the model can be used to help select efficient pulse sequence parameters when setting up an in vivo arterial inversion experiment. In addition, the model provides valuable insights into the adiabatic behavior of arterial spins. These insights could be useful in selecting an efficient surface coil geometry which achieves maximum inversion with a minimum B1 magnitude. © 1997 John Wiley & Sons, Ltd.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.