Abstract

We describe an infrequent case of endoleak Type-III due to an unrecognized, inadvertent bilateral limb deployment into the same limb gate of the Ovation aortic endograft, accompanied by thrombosis and acute ischemia. The following computed tomography angiography revealed the open limb gate with the characteristic of radio-opaque polymer in the sealing rings. Intraoperative angiographies via the brachial route identified the open limb gate and facilitated the successful use of an occluding plug to manage the Type-III endoleak.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.