Abstract

Photoacoustic (PA) effect has been widely applied in many fields, e.g., physics, chemistry and biomedicine. Herein, a miniaturized PA device is developed by integrating laser source, photo chopper, PA cell, microphone, and laptop for point-of-care testing in bioassay. With glucose assay as model, a piece of paper strip preloading chitosan, starch-potassium iodide (KI) and glucose oxidase (GOD) as lab-on-paper is employed for loading sample prior to PA detection. In the presence of glucose, the product generated on the paper strip would give rise to a strong PA signal in the PA cell under the irradiation of frequency-modulated laser at 520 nm via laptop readout. With a sample volume of 20 μL, a detection limit of 0.03 mM is obtained for glucose assay, along with a linear range of 0.08–1 mM. The accuracy and practicability of the present PA device is well demonstrated by detecting glucose in whole blood. Differing from the conventional PA instrument, the present PA device is really small in bulk with competitive sensitivity and excellent stability, offering a promising tool for point-of-care testing in bioassay.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.