Abstract

A microfluidic-capillary-waveguide-coupled fiber-optic sensor was developed for colorimetric determination of hazardous nitrite based on the Griess-Ilosvay reaction. The sensor was modularly designed by use of a light-emitting diode as the light source, silica fiber as the light transmission element, and a capillary waveguide tube as the light reaction flow cell. With the light interacting with the azo dye generated by the Griess-Ilosvay reaction between nitrite and Griess reagents, nitrite could be determined by a colorimetric method according to Beer's law. By use of the inexpensive and micro-sized elements mentioned above, the sensor provided a new low-cost and portable method for in situ and online measurement of nitrite. The sensor had a wide linear range for nitrite from 0.02 to 1.8 mg L(-1) and a low detection limit of 7 μg L(-1) (3σ), with a relative standard deviation of 0.37% (n = 10). With a low reagent demand of 200 μL, a short response time of 6.24 s, and excellent selectivity, the sensor is environmentally friendly and has been applied to nitrite determination in different water samples. The results were compared with those obtained by conventional spectrophotometry and ion chromatography, indicating the sensor's potential for practical applications.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call