Abstract

We present a miniature prototype of a resonance photoacoustic cell with a banana-shaped internal cavity with a volume of less than ∼5 mm3. The cell design and the experimental setup used for testing operational characteristics of the prototype are described. The results of testing the prototype in the experiments on photoacoustic detection of radiation absorption by ammonia in a nitrogen flow by using a single-mode diode laser generating radiation near ∼1.53 μm are presented. The amplitude-frequency characteristic of the photoacoustic response induced by absorption of radiation in the presence and absence of ammonia is analyzed. The experiment revealed that sensitivity of gas sensing is limited only by the microphone noise. The measured noise-limited minimal detectable absorption was found to be about 8.48 × 10−8 cm−1 W Hz−1/2.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.