Abstract

Traumatic brain injury (TBI) is a great potential threat to people who deal with explosive devices. Protection from TBI has attracted more and more interest. Great efforts have been taken to the studies on the understanding of the propagation of the blast events and its effect on TBI. However, one of the biggest challenges is that the current available pressure sensors are not fast enough to capture the blast wave especially the transient period. This paper reports an ultrafast pressure sensor that could be very useful for analysis of the fast changing blast signal. The sensor is based on Fabry-Perot (FP) principle. It uses a 45o angle polished fiber sitting in a V-groove on a silicon chip. The endface of the angle polished fiber and the diaphragm which is lifted off on the side wall of the V-groove form the FP cavity. The sensor is very small and can be mounted on different locations of a helmet to measure blast pressure simultaneously. The tests were conducted at Natick Soldier Research, Development, and Engineering Center (NSRDEC) in Natick, MA. The sensors were mounted in a shock tube, side by side with the reference sensors, to measure a rapidly increased pressure. The results demonstrated that our sensors' responses agreed well with those from the electrical reference sensors and their response time is comparable.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call