Abstract

Diabetes is a chronic metabolic disease with a high blood glucose level, leading to both seriously acute and chronic complications. The closed-loop system is an ideal system for diabetes management. However, the large size and high cost of the commercial systems restrict their widespread uses. Here, we present for the first time a microtube-based wearable closed-loop minisystem for diabetes management. The closed-loop minisystem includes a biosensing device, an electroosmotic micropump, and a printed circuit board (PCB) with an algorithm. The microtube-based sensing device coated on the outer surface of the microtube is inserted into subcutaneous tissue for detecting interstitial glucose; the current signal for sensing glucose is processed by the PCB to power the electroosmotic micropump intelligently for the delivery of insulin into the subcutaneous tissue via the microtube channel. The closed-loop minisystem worn on a diabetic SD rat can successfully maintain its blood glucose level within a safe level. It is expected that this new closed-loop paradigm could open up new prospects for clinical diabetes management.