Abstract

ABSTRACTA micro-fabricated pore is constructed and tested so that it can be used to characterize biological entities. The pore is prepared by bulk micro-machining of a silicon wafer. An oxide coated silicon diaphragm with the pore is placed between two chambers containing ionic buffer solutions to mimic a bilayer system. If a voltage is applied across the pore, electrophoretic passage of charged entities can be electrically detected through changes in the ionic current flow. When the entities traverse the pore, the ionic current is blocked and a decrease in the current can be observed. As an initial test case, negatively charged polystyrene beads which were 2.38μm in diameter, were electrophoretically driven across the pore. Then the bacteriumListeria innocua, suspended in Tris-glycine buffer, was also electrophoretically driven through the pore and its effective mobility was extracted. The device can also be used to study the interactions between organisms and the micro-fabricated surfaces. Work is continuing to scale the pore to the sub-100A range to be used for characterization and possible sequencing of single molecules such as DNA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.