Abstract

Transdermal extraction of interstitial fluid (ISF) offers an attractive method for non-invasive blood glucose monitoring. In order to calculate blood glucose concentration accurately, precise volume measurement of transdermally extracted ISF is required due to human skin’s varying permeability. In this paper, we presented a novel flow sensor fabricated from polydimethylsiloxane (PDMS), designed to measure the volume of conductive liquid. The flow sensor consists of two pairs of metal electrodes, which are fabricated in the PDMS channel. The volume of liquid is measured utilizing the time-of-flight of the two electrode pairs’ resistance while the liquid is flowing through the flow sensor. 1–14 μL normal saline solution was measured, the flow sensor measured volumes correlate very well (R 2 = 0.9996 and R 2 = 0.9975 for vacuum pump and syringe pump situations respectively) with the actual volumes. And the coefficient of variation for 10 times 10 μL normal saline solution measurement is 0.0077 (vacuum pump) and 0.0381 (syringe pump), respectively. The demonstrated flow sensor provides excellent functionality for conductive liquid.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.