Abstract

A 5 mm diameter tri-axial force sensor has been developed for minimally invasive robotic surgery. To define the required force range and resolution, a needle driver has been equipped with strain gauges. In vivo tests with different types of needles and tissue show that the required force range and resolution are, respectively, 2.5 N and 0.01 N. The new sensor is based on a flexible titanium structure of which the deformations are measured through reflective measurements with three optical fibres. It has a range of 2.5 N in axial direction and 1.7 N in radial direction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.