Abstract
Liquid scintillation counter (LSC), with its advanced spectrometry capabilities, the Quantulus 1220™, was used to determine gross α-β measurements in 25 environmental water samples. The determination of gross α-β emitters with the LSC method requires an accurate pulse shape analyzer (PSA) setting under various quenching conditions. External quenching parameter values were evaluated with americium and strontium standard solutions to look for the optimal PSA value. A PSA calibration curve was obtained and used to measure real samples. To validate a constructed PSA calibration curve, a test was performed using spiked samples with different standard solutions. The results of the same set of 25 water samples were verified by element specific separations followed by α-spectrometry. An acceptable correlation was obtained between the two techniques. In the case of LSC analysis, the results indicate a good PSA optimization. The method was found to be reliable, effective and very suitable for wide range measuring campaigns.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.