Abstract

To develop an image analysis method for calculation of angular disparities between an object, temporarily equipped with a reference system, and a radiographic receptor. A mathematical method based on a reference system containing two metallic spheres is developed, allowing calculation of inclination between the inter-spherical axis and the digital image receptor using image features. Experimental evaluation was done in standardized projection geometry for two sphere sizes at four randomly chosen inclinations per size, with each radiograph assessed three times. Truth was assessed threefold from photographs obtained at each inclination. Mean standard deviation between single assessments was 2.6 degrees. Significant differences (P Maloney/Rastogy=0.00) were found between absolute values of truth and calculated values (mean: -0.9 degrees; range: -6.0 degrees; 3.6 degrees), indicating a significant lack of accuracy. Although so far not sufficiently accurate, the method yields information relevant for correction of distortion in intra-oral radiology.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.