Abstract

A method is described that allows the measurement of the absolute value of the average energy to produce an ion pair, w, in noble gases for soft X-rays and its energy dependence. It uses a specially designed gas proportional scintillation counter working under electric fields below the ionization threshold. The spectrum of the radiation absorbed, and so the number of photons detected at each energy for a period of time on the order of the minute, is measured with the scintillation counter and, at the same time, the total primary electron charge produced is collected in one of the grids and measured with an electrometer. For pure xenon at 1030 mbar, a value of w=24.2+or-2.5 eV is measured with 5.9-keV X-rays. >

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.